Modeling the pulmonary mechanics of jet ventilation
Breidenstein, Max
Breidenstein, Max
Citations
Altmetric:
License
License
Abstract
Low frequency jet ventilation (LFJV) is an inadequately characterized technique to ventilate patients during rigid bronchoscopy. We conducted a benchtop experiment to characterize airway pressures and tidal volumes during LFJV under different levels of resistance, compliance, inspiratory time, and expiratory time. Airway pressures ranged from 7.69 – 12.02 cm H2O, median 10.99 cm H2O. Resistance and compliance had statistically significant but not clinically impactful effects on airway pressure. Tidal volumes ranged from 42.72 – 546.26 mL, median 177.00 mL. Resistance and compliance each had a statistically significant on tidal volume, but only compliance had a clinically relevant effect.
Description
Date
2025-06-02
Student Status
Medical Students
Journal Title
Journal ISSN
Volume Title
Type of presentation
Poster Presentation
Collections
Research Projects
Organizational Units
Journal Issue
Citation
DOI
Advisor(s)
Department
Program/Major
Medical Student
College/School
Larner College of Medicine
Organization
item.page.researchcategory
Clinical